4 Operation

792 Basic IC

70

Ident

User defined identifier (title) for the chroma-

togram to be displayed in the title bar of the

chromatogram window and in the

Chromatogram

open

window.

Calibration level

Calibration level (0 = sample; 1}n = calibra-

tion solutions).

Info 1 / Info 2

Sample

description.

Volume

Injected volume in PL.

Dilution

Dilution of the sample.

Vial number

Autosampler vial position to take sample from.

Amount

Sample amount. If this value is different for the

calibration run (c) and the sample run (s), the

component concentrations of the sample are

calculated as follows:

C

s

= C

c

Amount

s

/

Amount

c

Internal standard amount

Concentration of the internal standard compo-

nent for relative concentration calculations.

Date/time when...

Date and time of sample collection (the default

values are equal to the date and time when the

chromatogram starts).

SYSTEM / Control / Stop determination

Stop running determination. Data acquisition and time program are

terminated immediately. The recorded chromatogram is saved auto-

matically if the

Save chromatogram after the run

option on the

Passport /

Processing

tab is enabled.

Alternatively the determination can be stopped by clicking the icon

of the chromatogram window. In this case, the user is asked always if

the determination should be saved or not.

Print system parameters

SYSTEM / System / Print parameter

A report of the system parameters including the time program is cre-

ated and opened using the «Microsoft Word» program. The

*.rtf

file

opened can be printed, saved and exported into other programs. The

system report includes the name of the method linked to the system,

the configuration settings for the 792 Basic IC, the system startup val-

ues and the time program (incl.

Remote configuration

), if such a program

exists and is switched on (

ENABLED

).