3 Operating tutorial

792 Basic IC

48

6 Start determination

Ÿ Click on

<Start>

in the system window or select

Start determina-

tion

of the

Control

menu. A chromatogram window opens

where the baseline is recorded continuously. The status bar

of this window shows the message

Measure(Baseline)

, beside

this running time, analysis time, absolute conductivity and

number of measuring points per second are displayed. The

message

Waiting for INJECT [792COM#]

appears in the

SYSTEM

STATE

window.

7 Enter information for determination

Ÿ Enter the desired information concerning the sample in the

automatically opened

Edit

sample

description

window. Set

Cali-

bration level

to

0

and confirm with

<OK>

.

8 Condition system

Ÿ Let the system run for several minutes. If the system has

already been conditioned then a stable conductivity value of

approx. 16 PS/cm will be obtained after a few minutes.

Ÿ Double-click the chromatogram or select

View all

of the

View

menu. The sensitivity is set automatically so that all measur-

ing points are visible.

This function is only available if the integration has been

started. In the present example the integration only starts

after a delay period

Delay = 3.3 min

.

Ÿ Select the desired sensitivity using the cursor keys <Ï> or

<Ð>. In the chromatogram window the following display is

seen, for example: