4 Operation

792 Basic IC

74

Valve

Injection

valve

<Inject>

Switch injection valve to "INJECT" position.

<Fill>

Switch injection valve to "FILL" position

Suppressor

Suppressor

module

<Step>

Switch the suppressor module to the next posi-

tion. The time since the last switching of the

suppressor module is displayed in the field be-

side the

<Step>

button.

IC pump

High-pressure

pump

<On>

Start high-pressure pump.

<Off>

Stop high-pressure pump.

Peristaltic

Peristaltic

pump

<On>

Start peristaltic pump.

<Off>

Stop peristaltic pump.

System startup values

The system startup values are sent and applied

to the 792 Basic IC each time the system is

connected, a determination is started, or the

values are sent manually with

<Send to unit>

.

Flow, mL/min

Startup value for flow rate of the high-pressure

pump.

Entry range:

0.20

...

2.50 mL/min

Pressure max, MPa

Startup value for maximum pressure limit for

high-pressure pump. This limit is controlled

even without connection to the PC.

Entry range:

0.0

...

25.0 MPa

Pressure min, MPa

Startup value for minimum pressure limit for

high-pressure pump. This limit is controlled

even without connection to the PC.

Entry range:

0.0

...

25.0 MPa

Time program

On the

Program

tab of the system settings window a user-defined time

program for instrument control can be entered. This program is started

automatically at the moment the determination is started. The time pro-

gram contains program steps including time, program instruction and

parameter.