5.2 Maintenance and servicing

792 Basic IC

149

6 Mounting the pump head

x Reassemble the components of the pump head 34

34

34

34 as shown

in Fig. 16. Tighten the screw 81

81

81

81 by hand. First screw in piston

cartridge 85

85

85

85 manually until the stop is reached and then use

a wrench to turn it through a further

15°. Firmly tighten the two

valve screw holders 92

92

92

92 with a wrench.

x Reattach pump head 34

34

34

34 to the pump with the help of the 4

fixing screws 33

33

33

33. Firmly tighten them with the 6.2621.030 Al-

len key.

To ensure that the pump head is not positioned wrongly, the holes at

the rear for the clamping bolts have different depths, i.e. 1 clamping

bolt is longer than the rest. The deepest hole must naturally accom-

modate the longest bolt. If this is not the case, the pump will not

function properly.

x Screw connection capillary 35

35

35

35 to pump head 34

34

34

34 (see Fig. 3).

x Connect aspirating tubing to aspirating capillary 32

32

32

32 at the

pump head 34

34

34

34.

Inlet valve

Outlet valve

6.2824.090 6.2824.080

95

95

95

95

96

96

96

96

97

97

97

97

98

98

98

98

99

99

99

99

100

100

100

100

101

101

101

101

95

95

95

95

96

96

96

96

97

97

97

97

100

100

100

100

99

99

99

99

98

98

98

98

101

101

101

101

Fig. 18: Components of inlet valve 90 and

outlet valve 91

95

95

95

95

Valve housing

96

96

96

96

Sealing ring (black)

97

97

97

97 Sleeve

98

98

98

98 Sapphire sleeve

The bright side must point towards

the sapphire sphere

99

99

99

99 Sapphire sphere

100

100

100

100 Ceramic holder for sapphire

sphere

The large recess must point in the

direction of the sapphire sphere

101

101

101

101 Seal

The larger opening must point

outwards