Metrohm 792 Basic IC User Manual

Page 111

Advertising
background image

4 Operation

792 Basic IC

102

Calibration curve

792 BASIC IC / Method / Calibration / Graphs

Selecting this menu item or clicking on

<Graphs>

in the component table

opens the

Component

window.

In this window the results of the calibration are shown for each compo-

nent together with the calibration curve. The parameters for calculating

the calibration curve can also be entered here.

Calibration results and curve

Analytical expression

A line above the calibration curve looking like
Q

=

=

=

= k

3

xxxxA

3

+ k

2

xxxxA

2

+ k

1

xxxxA + k

0

.

This formula is used for approximation of the

calibration curve.

RSD

RSD (Residual Standard Deviation) value to

evaluate the error of calibration curve approxi-

mation.

Corr.

Correlation coefficient value is available only in

the case of linear calibrations without weighting

of data points.

Calibration curve

Plot of measured calibration points and calcu-

lated calibration curve.

k0, k1, k2, k3

Calibration coefficients k

0

, k

1

, k

2

and k

3

(coeffi-

cients of the calibration formula).