Metrohm 792 Basic IC User Manual

Page 119

Advertising
background image

4 Operation

792 Basic IC

110

Channel table

Parameters of the data acquisition channel

used.

CHANNELS TABLE

No Name Units Input Minimum Zero Maximum Range Coefficient Noise Shift

1 Cond uS/cm 1 -8388607 0 8388607 8.389e+01 1.000e-05 1.52 0

# Name/Units Noise RMS PeakToPeak Drift/hour

1 Cond 1.5 52871.6 512493.0 4232.464

uS/cm 1.52e-05 0.529 5.12 0.0423

Name

Channel’s name. This text appears as

Channel

label

in the chromatogram.

Units

Units of detector response (PS/cm).

Input

Number of data acquisition channel.

Minimum

Minimum value of the linear range of the AD

converter (in ADC conversion units). This pa-

rameter is used to detect an underflow condi-

tion.

Zero

A signal level on a baseline (in ADC conversion

units). This parameter is a result of the ADC

calibration.

Maximum

Maximum value of the linear range of the AD in-

terface (in ADC conversion units). This parame-

ter is used to detect overflow condition.

Range

An input signal value in maximum in PS/cm.

Range

= (

Maximum

-

Zero

)

Coef

Coef

ADC sensitivity coefficient (weight of ADC bit in
PS/cm).

Noise

Estimated baseline noise value of the channel

in ADC conversion units (bits).

Shift

Not used for 792 Basic IC.

Report destination
The following output devices can be selected simultaneously in any

combination as targets for report output:

Screen

Output report (without curves) to screen.

Printer

Output report to printer.

File

Output report to file (see File output options).

Advertising