Percentage calibration

Setup value can be configured at the ratio to
the current received light intensity.

The percentage calibration target value can be
set in the range of "-99P" (-99%) to 99P (99%)
Default: -10P (-10%)



Sensitivity setting

1

In the desired reference state, press
the [SET] button.
Calibration is complete after the setting
value flashes momentarily, and then
stops (lights up).

• If the external input setting is set to "

SEt

" (external calibration input), peri-

odic percentage calibration is possible from external devices, enabling
stable detection even on workpieces having small sensitivity differences.

"External Input" (page 4-17)

• If the detection mode is set to "

ArEA

(area detection mode)", Hi and Lo

are set simultaneously according to the received light intensity.

(Example)
If the percentage calibration is set
to 10P (10%) when the reference
received light intensity is "100",
the Hi and Lo setting values will be
set as follows.
Hi setting value: 110
Lo setting value: 90

900

1000

Reference

Received light intensity

Setting example

When the percentage calibration
target value is set to "-10%".

Reference

Received

light intensity

Hi

110

100

90

Lo