Calibration

97

Analog Input Module‘s Calibration Voltages

Table 5-1: ND-6011/D/ND-6018 Calibration voltages

Code Input

Range

Offset Calibration

voltage

Span Calibration

Voltage

00

±15 mV

0 mV

+15 mV

01

±50 mV

0 mV

+50 mV

02

±100 mV

0 mV

+100 mV

03

±500 mV

0 mV

+500 mV

04

±1 V

0 V

+1 V

05

±2.5 V

0 V

+2.5 V

06

±20 mA

0 mA

+20 mA

0E

J (0

°C to 760°C)

0 mV

+50 mV

0F

K (0

°C to 1000°C)

0 mV

+55 mV

10

T (-100

°C to 400°C)

0 mV

+30 mV

11

E (0

°C to 1000°C)

0 mV

+78 mV

12

R (500

°C to 1750°C)

0 mV

+25 mV

13

S (500

°C to 1750°C)

0 mV

+19.5 mV

14

B (500

°C to 1800°C)

0 mV

+15 mV

15

N (-270

°C to 1300°C)

0 mV

+55 mV

16

C (0

°C to 2320°C)

0 mV

+39 mV

Table 5-1: ND-6012/D/ND-6017 Calibration voltages

Code Input

Range

Offset Calibration

voltage

Span Calibration

Voltage

08

±10 V

0 mV

+10 V

09

±5 V

0 mV

+5 V

0A

±1 V

0 mV

+1 V

0B

±500 mV

0 mV

+500 mV

0C

±150 mV

0 mV

+150 mV

0D

±20 mA

0 mA

+20 mA