14 offset calibration to each channel – ADLINK ND-6018 User Manual

Page 59

Advertising
background image

Command Set

51

3.14 Offset Calibration to each Channel

@Description

To correct the offset errors of AD converter by using the offset
calibration.

@Syntax

$(Addr)1(Channel No)<CR>

$

Command leading code

(Addr)

Address ID

1

Offset calibration.

(Channel No)

Channel for calibration.(1 character)0~2

@Response

!(Addr)<CR>
or
?(Addr)<CR
>

!

Command is valid.

?

Command is invalid.

(Addr)

Address ID.

@Example

User command:

$0612<CR>

Response:

!06<CR>


To perform the offset calibration for analog input module ND-6013
channel 2 at specified address 06 (Hex).

Note:

To perform the calibration, a proper input signal should be
connected to the analog input module. Different input range
have different input voltage, detail refer chapter 5
Calibration” .

(6013)

(for 6013 F/W version
A3.05~A4.60)

Advertising