Introduction, Electrical safety analyzer – Fluke Biomedical ESA612 User Manual

Page 15

Advertising
background image

1

Electrical Safety Analyzer

Introduction

The Fluke Biomedical ESA612 Electrical Safety Analyzer
(hereafter the Analyzer) is a full-featured, compact,
portable analyzer, designed to verify the electrical safety
of medical devices. The Analyzer tests to domestic
(ANSI/AAMI ES1, NFPA 99) and international (IEC62353,
AN/NZS 3551, and parts of IEC 60601-1) electrical-safety
standards. The integrated ANSI/AAMI ES1 and
IEC60601-1 patient loads are easily selectable.
The Analyzer performs the following tests:

•

Line (Mains) voltage

•

Ground Wire (or Protective Earth) Resistance

•

Equipment

current

•

Insulation

resistance

•

Ground (Earth) leakage

•

Chassis (Enclosure) leakage

•

Lead to Ground (Patient) and Lead to Lead (Patient
Auxiliary) leakage

•

Lead isolation (Mains on applied parts leakage)

•

Differential

leakage

•

Direct equipment leakage

•

Direct applied part leakage

•

Alternative equipment leakage

•

Alternative applied part patient leakage

•

Point to point leakage, voltage, and resistance

•

ECG simulation and performance waveforms