Fluke Biomedical ESA612 User Manual

Page 70

Advertising
background image

ESA612
Users Manual

56

When performing an applied parts test using the
AAMI/NFPA-99 standard, the normal connections of RA,
LL, LA, and RL are made to their associated Analyzer
input jacks. Four adapters from the Universal Snap to
Banana Adapter set will be required for the first four
connections. The remaining chest leads are connected to
the Adapter and the common lead from the Adapter is
plugged into the V1 jack (5

th

jack) of the Analyzer. See

Figure 31. This configuration allows for isolation of the RA,
LL, LA, and RL leads from each other and the remaining
chest leads, which are shorted together, while performing
leakage tests.