Fluke Biomedical ESA612 User Manual

Page 88

Advertising
background image

ESA612
Users Manual

74

Differential leakage

Ranges ................................................................ 75 to 199

μ

A

200 to 1999

μ

A

2.00 to 20.00 mA

Accuracy .............................................................

±

(10 % of reading + (2 counts or 20

μ

A, whichever is greater))

Insulation resistance

Ranges ................................................................ 0.5 to 20.0 M

Ω

20.0 to 100.0 M

Ω

Accuracy

20 M

Ω

Range ..................................................

±

(2 % of reading + 0.2 M

Ω

)

100 M

Ω

Range ................................................

±

(7.5 % of reading + 0.2 M

Ω

)

Source test voltage ............................................. 500 V dc (+20 %, -0 %) 2.0

±

0.25 mA short-circuit current or 250 V dc selectable

Maximum load capacitance ................................ 1

μ

F

ECG Performance Waveforms

Accuracy

Frequency .......................................................

±

2 %

Amplitude ........................................................

±

5 % of 2 Hz square wave only, fixed @ 1 mV Lead II configuration

Waveforms

ECG Complex ................................................. 30, 60, 120, 180, and 240 BPM
Ventricular Fibrillation
Square wave (50 % duty cycle) ...................... 0.125 Hz and 2 Hz
Sine wave ....................................................... 10, 40, 50, 60, and 100 Hz
Triangle wave .................................................. 2 Hz
Pulse (63 ms pulse width) ............................... 30 BPM and 60 BPM