Performing a direct equipment leakage test – Fluke Biomedical ESA612 User Manual

Page 60

Advertising
background image

ESA612
Users Manual

46

To perform an alternative applied part leakage test:
1. Press

.

2. Press the soft key labeled More.
3. Select the desired applied part groupings using

and .

4. Press the soft key labeled Select.
5. Press the soft key labeled Alternative A.P..
6. Press

to apply the test voltage and read the

current in the display.

7. Press

or  to advance to the next applied part

group(s) of a single function if applicable. Pressing

to read leakage current for each group.

Figure 26 shows the electrical connections between the
Analyzer and the DUT during an Alternative Applied Part
Leakage current test.

Note

If there are more than five applied parts to
connect to the Analyzer, see Using the 1-to-10
Adapter later in this manual.

Performing a Direct Equipment Leakage Test

Note

The Direct Equipment Leakage test is available
when the EN62353 standard is selected.

The Direct Equipment Leakage Current test measures the
leakage current between all applied parts and the exposed
conductive surface on the housing, to mains earth.
To perform a direct equipment test:
1. Press

.

The direct equipment test is the default test and should
already be selected.
2. Press

to apply the voltage and read the leakage

current in the display

Figure 27 shows the electrical connections between the
Analyzer and the DUT during a Direct Equipment Leakage
Current Test.