Deleting data from memory, Controlling the analyzer remotely – Fluke Biomedical ESA612 User Manual

Page 78

Advertising
background image

ESA612
Users Manual

64

Deleting Data from Memory

To delete a Test Record and its associated data from
memory:
1. Press

.

2. Press the softkey labeled More to reveal additional

menu selections.

3. Press the softkey labeled View Memory.
4. Select the desired Test Record by pressing  or 

to scroll through the records listed in the display.

5. Press the softkey labeled Delete.

Note

All Test Records can be deleted at this point by
pressing the softkey labeled Delete All.

6. When

Delete?

appears in the display, press the

softkey labeled Delete to remove the record from
memory. Otherwise press the softkey labeled Cancel
to return to the Test Record list and leave the
selected record in memory.

Note

Deleting the last or current record does not open
the previous record for additional data storage.
Only a new Test Record can be opened for data
recording after deleting the last record.

Controlling the Analyzer Remotely

Fluke Biomedical Ansur test automation software allows a
solutions-based approach to complete testing of the
medical device under test (DUT). Ansur helps create
standard work using the test template/sequence (which is
based on a user written test procedure), and integrates all
test results into a single test report which can be printed or
archived. Ansur allows for automatic comparisons to the
limits of the standard selected, indicating whether results
are passing or failing. Ansur manages test procedures by
allowing both manual and visual automated test
sequences.

Advertising