Intended use – Fluke Biomedical ESA612 User Manual

Page 18

Advertising
background image

ESA612
Users Manual

4

The Analyzer must be properly earthed.

Only use a supply socket that has a
protective earth contact. If there is any
doubt as to the effectiveness of the supply
socket earth, do not connect the Analyzer.
Do not use a two-conductor adapter or
extension cord; this will break the
protective ground connection.

Do not use the 15-20 A adapter to power

devices rated in excess of 15 A. Doing so
may overload the installation.

Use extreme caution when working with

voltages above 30 V.

Use the proper terminals, functions and

ranges for the test being performed.

Do not touch metal parts of the device

under test (DUT) during analysis. The DUT
should be considered an electrical shock
hazard when connected to the Analyzer as
some tests involve high voltages, high
currents, and/or the removal of DUT earth
bond.

Intended Use

The Product is an electronic signal source and
measurement device for verifying the electrical safety of
medical devices. The Product also provides ECG
simulation and performance waveforms to verify patient
monitors are performing within their operating
specifications.
The Product provides the following function categories:

ECG

Functions

ECG-Performance

Testing