Figure 6 shows polarity is settable between normal,
reversed, and off. Neutral is also settable to closed or
open. Earth condition is not displayed, which indicates it
can not be changed. However, earth is internally opened
during this test.

Setting Up the Analyzer

There are a number of Analyzer parameters that are
adjusted through a setup function as well as the ability to
save a record by ID and date. To access the first Setup
menu shown in Figure 7, press .

fis114.bmp

Figure 7. Setup Menu

Note

See the Using Memory section later in this
manual for a description of how to enter a test
record ID.

The setup parameters have been grouped into six
categories: Instrument, Display, Sound, Instrument Info,
Calibration, and Diagnostics.

Setting Polarity Switching Delay

When switching the polarity of the Analyzer’s test
receptacle, a delay can be set to control the actual switch
time. To set the polarity delay:
1. From the setup menu, press the softkey labeled

More

to reveal additional menu selections.

2. Press the softkey labeled Instrument to reveal the

instrument setup selections.

3. Press the softkey labeled Polarity Delay to open the

scroll box above the softkey label.

4. Press

or  to adjust the delay from 1 to 5

seconds in 1 second steps.