Electrical Safety Analyzer

Safety Information

3

Safety Information

In this manual, a Warning identifies hazardous conditions
and actions that could cause bodily harm or death. A
Caution

identifies conditions and actions that could

damage the Analyzer, the equipment under test, or cause
permanent loss of data.

 Warning

To avoid possible electrical shock or
personal injury, follow these guidelines:

•

Use this Analyzer only in the manner

specified by the manufacturer or the
protection provided may be impaired.

•

Read the Users Manual before operating

the Analyzer.

•

Do not connect the Analyzer to a patient or

equipment connected to a patient. The
Analyzer is intended for equipment
evaluation only and should never be used
in diagnostics, treatment or in any other
capacity where the Analyzer would come
in contact with a patient.

•

Do not use the product in wet or damp

locations, around explosive gases or dust.

•

Inspect the Analyzer before using it. Do

not use the Analyzer if abnormal
conditions of any sort are noted (such as a
faulty display, broken case, etc.)

•

Inspect the test leads for damaged

insulation or exposed metal. Check test
lead continuity. Replace damaged leads
before using the Analyzer.

•

When testing, always be sure to keep your

fingers behind the safety barriers on the
test leads.

•

Never open the Analyzer's case.

Dangerous voltages are present. There are
no user replaceable parts in the Analyzer.

•

Have the Analyzer serviced only by

qualified personnel.