Viewing memory, Setting the gfci limit, Performing electrical safety tests – Fluke Biomedical ESA612 User Manual

Page 30: Setting the test standard

Advertising
background image

ESA612
Users Manual

16

Viewing Memory

Refer to the Using Memory section later in the manual to
learn about viewing memory and how to store data in the
Analyzer.

Setting the GFCI Limit

To set the GFCI current limit:
1. From the setup menu, press the softkey labeled

More

to reveal additional menu selections.

2. Press the softkey labeled Instrument to reveal the

instrument setup selections.

3. Press the softkey labeled GFCI Limit to open the

scroll box above the softkey label.

4. Press

or  to adjust the current limit from 5 to

25 mA.

5. Press the softkey labeled Done to exit the GFCI Limit

setup function.

Performing Electrical Safety Tests

The Analyzer is designed to perform a number of different
electrical and performance tests on biomedical
equipment. The following sections describe the various
tests and how to perform them using the Analyzer.

Setting the Test Standard

The Analyzer is designed to perform electrical safety
testing based on a number of different safety standards:

AAMI ES1/NFPA99, IEC62353, IEC60601-1, and
AN/NZS 3551. AAMI is the Analyzer’s default standard.
To select another standard:
1. Press

.

2. From the setup menu, press the softkey labeled

More

to reveal additional menu selections.

3. Press the softkey labeled Instrument to reveal the

instrument setup selections.

4. Press the softkey labeled Standard to open the

scroll box above the softkey label.

5. Press

or  to scroll through the standard

selections.