Fluke Biomedical ESA612 User Manual
Page 45
Advertising

Electrical Safety Analyzer
Performing Electrical Safety Tests
31
The following outlet conditions apply when performing this
test:
•
Normal
Polarity
•
Normal Polarity, Open Neutral
•
Reversed
Polarity
•
Reversed Polarity, Open Neutral
IEC60601-1 specifies that the applied parts should be
connected for this measurement. Enable this
measurement by pressing or which grounds and
ungrounds all applied parts connection posts.