ESA612
Users Manual

60

To select one of the predefined waveforms, press the
softkey labeled Wave Form. A scroll box with  next to it
appears above the softkey label. Use  or  to scroll
through the different waveforms.
For all waveforms except VFIB and Triangle, the rate or
frequency of the waveform is adjusted through the softkey
labeled Frequency or Rate. For some waveforms, there
are more than two frequency or rate selections. For those
waveforms, pressing the softkey labeled Frequency or
Rate

will open a scroll box above the softkey label with 

next to it. Use  or  to select the frequency or rate. For
those waveforms that have only two frequency or rate
selections, the softkey labeled Frequency or Rate acts as
a toggle, where each press of the softkey switches to the
other value.