Performing a current consumption test, Performing leakage current tests – Fluke Biomedical ESA612 User Manual

Page 43

Advertising
background image

Electrical Safety Analyzer

Performing Electrical Safety Tests

29

Performing a Current Consumption Test

To measure the current consumed by the DUT, press

. The Analyzer displays the current flowing through

the mains connections of the test receptacle.

Performing Leakage Current Tests

The Analyzer measures leakage current for a number of
different DUT configurations. In addition to the leakage
found on the enclosure and the earth connection, the
Analyzer can measure leakage on each connected applied
part and combinations of connected applied parts.

Which leakage tests are available depends on which
standard is selected. See the Selecting the Test Standard
section earlier in this manual to change the standard the
Analyzer is using.
Table 5 lists six leakage current tests that have different
names based on which standard is selected.
Press  to access the leakage current main menu
shown in Figure 18.

Table 5. Test Names Based on Selected Standard

IEC60601 AAMI/NFPA

99

Protective Earth Resistance

Ground Wire Resistance

Earth Leakage Current

Ground Wire Leakage Current

Touch or Enclosure Leakage Current

Chassis Leakage Current

Patient Leakage Current

Lead to Ground Leakage Current

Patient Auxiliary Leakage Current

Lead to Lead Leakage Current

Mains on Applied Part (MAP) Leakage Current

Isolation Leakage Current