Fluke Biomedical ESA612 User Manual

Page 87

Advertising
background image

Electrical Safety Analyzer

Detailed Specifications

73

Crest factor ..........................................................

3

Ranges ................................................................ 0.0 to 199.9

μ

A

200 to 1999

μ

A

2.00 to 10.00 mA

Accuracy

DC to 1 kHz .....................................................

±

(1 % of reading + (1

μ

A or 1 LSD, whichever is greater))

1 to 100 kHz ....................................................

±

(2 % of reading + (1

μ

A or 1 LSD, whichever is greater))

1 to 5 kHz (current > 1.6 mA) ..........................

±

(4 % of reading + (1 μA or 1 LSD, whichever is greater))

100 kHz to 1 MHz ............................................

±

(5 % of reading + (1

μ

A or 1 LSD, whichever is greater))

Note

Accuracy for Isolation, MAP, Direct AP, Alternative AP, and Alternative Equipment leakage tests all ranges are:

At 120 VAC + (2.5 µA or 1 LSD, whichever is greater)

At 230 VAC additional ± 3.0 % and + (2.5 µA or 1 LSD, whichever is greater)

For Alternative and Direct AP leakage tests, the leakage values are compensated for nominal mains as per 62353.
Therefore, the accuracy specified for other leakages is not applicable.

Mains on applied part test voltage ...................... 100 %

±

7 % of Mains for AAMI, current limited to 1 mA

±

25 % per AAMI

100 %

±

7 % of Mains for IEC 62353 current limited to 3.5 mA

±

25 % per IEC 62353

100 %

±

7 % of Mains for IEC 60601-1 current limited to 7.5 mA

±

25 % per IEC 60601-1

Advertising