background image

37

Place restorations on the honey-comb firing tray and fire.

Build up the materials in accordance with the reduction.

For a life-like incisal third, combine various materials and layer.

Firing parameters for the Wash firing

(note the temperature control)

IPS Empress CAD with

B

S

t

T H

V

1

V

2

IPS Empress Esthetic Veneer

Wash firing

403°C

4'

60°C

840°C

2'

450°C

839°C

757°F

4'

108°F

1544°F

2'

842°F

1543°F

Firing parameters for the 1

st

Incisal/Transpa firing

(note the temperature control)

IPS Empress CAD with

B

S

t

T H

V

1

V

2

IPS Empress Esthetic Veneer

1

st

Incisal/Transpa firing

403°C

4'

60°C

830°C

2'

450°C

829°C

757°F

4'

108°F

1526°F

2'

842°F

1525°F

1

st

Incisal/Transpa firing with IPS Empress Esthetic Veneer

The incisal third is built-up according to the silicone key. The matching layering materials are listed in the Combination

Table on Page 44. The following working procedure should be observed:

– Mix IPS Empress Esthetic Veneer ceramic materials with the IPS Empress Esthetic Veneer Build-Up Liquid.

– The mixed material should demonstrate good stability. If this is not the case, either too much or too little Build-Up Liquid

was added.

– Build up the materials in accordance with the reduction using a ceramic brush.

– If a life-like incisal third is desired, various materials maybe combined and layered.

– Once the tooth shape has been completed, allow the ceramic material to dry for a short period of time.

– Place veneers and anterior crowns on the honey-comb firing tray with the help of metal pins. Additionally support the

veneers with a small amount of IPS Object Fix.

B = Stand-by temperature °C / °F
S = Closing time/minutes
t

= Temperature increase rate °C/min. / °F/min.

T = Firing temperature °C / °F
H = Holding time / min.

V

1

= Vacuum-on temperature °C / °F

V

2

= Vacuum-off temperature °C / °F

B = Stand-by temperature °C / °F
S = Closing time/minutes
t

= Temperature increase rate °C/min. / °F/min.

T = Firing temperature °C / °F
H = Holding time / min.

V

1

= Vacuum-on temperature °C / °F

V

2

= Vacuum-off temperature °C / °F