Ivoclar Vivadent IPS Empress CAD Chairside - de-en-es-it-ru User Manual

Page 88

Advertising
background image

40

Optional

Subsequent adjustment

After completion, additional adjustment (e.g. contact points) may become necessary. For that purpose,

IPS Empress Add-On 770 °C/1418 °F is available.

Procedure:

– Before the adjustment, the restoration must be free from dirt and grease. For that purpose, thoroughly clean the

restoration using the steam jet.

– Mix IPS Empress Add-On 770 °C / 1418 °F to a workable consistency. Make sure that the add-on material and liquid are

evenly mixed so that an optimum firing result can be achieved.

– Apply the mixed add-on material on the cleaned, desired areas using a brush and slightly blot them with an absorbent

cloth.

– Position the adjusted restoration on the firing tray and fire in the ceramic furnace.

– After firing, manually polish the supplemented areas to a high gloss.

IPS Empress CAD with

B

S

t

T H

V

1

V

2

IPS Empress Add-On 770°C/1418°F

Corrective firing

403°C

4'

60°C

770°C

2'

450°C

769°C

757°F

4'

108°F

1418°F

2'

842°F

1417°F

Firing parameters for the corrective firing

IPS Empress CAD cut-back and layered – the ideal combination for highly aesthetic, CAD/CAM fabricated restorations.

B = Stand-by temperature °C / °F
S = Closing time/minutes
t

= Temperature increase rate °C/min. / °F/min.

T = Firing temperature °C / °F
H = Holding time / min.

V

1

= Vacuum-on temperature °C / °F

V

2

= Vacuum-off temperature °C / °F