8 Good pump installation practice

8.1 General recommendations

Always consult an expert before installing a metering pump in a

specialised system. Metering pumps should be maintained by qualified

persons.

It is recommended that the pump is bolted to a flat, horizontal, rigid surface, free

from excessive vibration, to ensure correct lubrication of the gearbox and correct

pumphead operation. Allow a free flow of air around the pump to ensure that heat

can be dissipated. Ensure that the ambient temperature around the pump does not

exceed 45C (113F).

The STOP key on pumps supplied with a keypad will always stop the pump. However,

it is recommended that a suitable local emergency stop device is fitted into the

mains supply to the pump.

Do not stack pumps.

This pump is self-priming and self-sealing against backflow. No valves are required

in inlet or discharge lines, except as described below. Valves in the process flow must

be opened before the pump operates.

Users must fit a non-return valve between the pump and the

discharge pipework to avoid the sudden release of pressurised

fluid in the event of a pumphead failure. This shall be fitted

immediately after the discharge of the pump.

qdos30 is a positive displacement pump; therefore, it is recommended

best practice that customers incorporate discharge pressure relief in their

piping system. Failure to fit a pressure release valve in the discharge

pipework will result in excessive build up of pressure should the discharge

become blocked. This may present a safety risk, may cause damage to

the system pipework or lead to premature failure of the pumphead. The

pressure relief valve shall be rated at no more than 10 bar. It shall always

be rated at below the users system maximum operating temperature. It

shall be installed so as to provide easy access for maintenance, inspection

and repair. It shall not be capable of being adjusted without the use of

a tool. The discharge opening must be located and directed so that the

release material is not directed towards any person and will not deposit

on parts that could cause a hazard. There must not be a shut-off valve

between the overpressure safety device and the pump.

Do not block the drain port of the ReNu pumphead.

Ensure the chemicals to be pumped are compatible with the

pumphead and the pipework and fittings to be used with the

pump. Please refer to the chemical compatibility guide which can

be found at: www.wmpg.com/chemical

If you need to use the pump with any other chemical please

contact Watson-Marlow to confirm compatibility.