15.3 General settings

To view the general settings menu, select GENERAL SETTINGS from the main menu.

Auto restart

Press ENABLE/DISABLE to turn the auto restart feature on/off

This pump offers an auto restart feature. If active on power loss, it restores the

pump when power returns to the operating state it was in when power was lost.

For example, if the pump was running in analog mode prior to power loss, it would

return to the same operating mode and continue to run at a proportional speed to

the analog input.

If power is lost in the middle of a dose, when the power is returned it will continue

and finish the interrupted dose.

Any pulses that were in memory before power loss will be remembered. Pulses

received during power loss will be lost.

Do not use auto restart for more than 20 starts per hour. We

recommend remote control where a high number of starts is

required.

The ! icon is displayed on the home screens to indicate that the auto restart feature

is active.

Flow units

The current chosen flow unit is displayed on the right hand side of the screen. To

change flow units move the selection bar over the flow unit menu entry and press

SELECT.