14.4 Manual mode operation when

PROFIBUS is enabled

In Manual mode, if PROFIBUS is enabled, the P symbol will appear in the icon area.

The P will be white if there is data exchange, it will be red if there is no data

exchange. If there is no data exchange, the INFO button will become BUS STATUS

in red.

Press BUS STATUS to get the PROFIBUS STATUS screen, the background will

indicate pump status. It will not be red because this is not a critical error. The pump

is still operating but is not reporting diagnostics to the master.

The MODE button gives access to the PROFIBUS settings and the station address.

When the menus are accessed, the pump is technically still in MANUAL mode but with

no PROFIBUS communications. After five minutes of inactivity the pump will revert

back to the MANUAL home screen and discard any unsaved changes, if there are still

no communications then the red P icon will be displayed. For further information on

PROFIBUS communication errors see section 14.7 PROFIBUS comunication errors.