Pressure Systems, Inc. Model 9116 User’s Manual

Page 97

www.PressureSystems.com

Following is a simple, step-by-step procedure for executing a specified-value span calibration of
a Model 9116 Intelligent Pressure Scanner. For the purposes of this example, it will be assumed
that an upscale pressure of 14.9800 psi is available from a dead weight tester for the calibration
of 15 psi internal transducers. All sixteen channels are 15 psi full scale.

Description

Model 9116 Command

Ensure that valves are in RUN/CAL mode (default)

[w1200]

… normal data acquisition

Perform Re-zero calibration first

See Section 4.2.2

Place the module calibration manifolds into the CAL position
if this is the desired pressure application input. The RUN
position may be a better choice for modules with
transducers having different ranges.

[w0C01] for CAL

[w0C00] for RUN

With a deadweight tester, apply 14.9800psi to the
appropriate module CAL and CAL REF inputs [or optionally
to RUN inputs]

Delay for settling of pneumatic inputs

Verify measured data reads near expected full scale

[rFFFF0]

Instruct module to calculate new gain coefficients for all 16
channels

ZFFFF 14.98

Place calibration manifold back into the RUN position

w0C00

Store new gain coefficients to transducer non-volatile
memory

w09

… continue normal data acquisition