Pressure Systems, Inc. Model 9116 User’s Manual

Page 33

www.PressureSystems.com

Example:

Configure and start the Multi-Point Calibration function so that it affects only the first

four (4) channels of the module. Three (3) pressure calibration points will by supplied

when we continue this function later (see example for ‘01’ sub-command below). A

linear (1st order) fit will be used to obtain a new set of offset and gain correction

coefficients for these four (4) channels. The maximum average sample count (32) is

used to collect each calibration data point, so as to minimize any noise in the data

samples. The module’s data acquisition process is altered immediately to collect the

increased number of averages.


C 00 F 3 1 32


Read

response:


A