Pressure Systems, Inc. Model 9116 User’s Manual

Page 74

www.PressureSystems.com

READ TRANSDUCER TEMPERATURE (Command ‘t’)


Purpose:

Returns the most recently acquired engineering-unit temperature data (in ºC) for

the specified channels. Each datum returned in the response will be in the

specified high-precision data format.


Command

tppppf

‘t’ is the command letter
pppp’ is the position field
f’ is the format field


Response

dddd.. dddd

dddd’ are the datum fields, each with leading space (except f =7
or 8).



Description: The 4-character hex position field (pppp) specifies a 16-bit binary bit-map, with

each set bit (1) specifying a particular channel number (16-1, left-to-right).

The

1-character

format field (f) specifies the format of each data field (dddd) that

will be returned in the requested response. The first datum returned in the
response will be for the highest channel number specified.
Each (non-
binary) datum will be preceded by a space character. Some formats may not be
applicable to the specific type of data being requested. Valid formats are shown
in the following table:

f

converts each internal response datum value from . . .

max. char.

0 single binary float

to 7-10-digit signed decimal “ [-xxx]x.xxxxxx”

13

1 single binary float

to 8-digit hex integer “ xxxxxxxx”

9

2 double binary float

to 16-digit hex integer “ xxxxxxxxxxxxxxxx”

17

5 single binary float

to long integer (EU*1000) then to 8-digit hex integer

9

7 single binary float

to single binary float (big endian: msb first)

4

8 single binary float

to single binary float (little endian: lsb first)

4