Measurement Specialties 9116 User Manual

Page 59

Advertising
background image

Pressure Systems, Inc. Model 9116 User’s Manual

Page 50

www.PressureSystems.com

Note

With the exception of binary formats 7 and 8, all other formats
include a leading space
in each datum delivered in each stream
packet.

The number of stream packets (‘num’) parameter is a positive integer count (from
0 to 2147483647, specified with 1 to 10 numeric digits as needed). It sets a finite
limit on the number of packets delivered in the host data stream. The value 0 for
this parameter requests “continuous” output packets for the defined host stream
(unbounded).

Note

While the ‘c 00’ command will allow hardware and software trigger
scan lists to operate concurrently, this is not a recommended mode
of operation. Running hardware and software trigger scan lists
concurrently diminishes the degree of determinism for the hardware
scan lists. The hardware trigger lists will determine the frequency of
pressure readings and the update of internal pressure, current value
table. The software trigger lists will return data from this table at the
requested software trigger rate, however the table will only be
updated at the rate of the fastest hardware trigger scan lists.

Advertising