Pressure Systems, Inc. Model 9116 User’s Manual

Page 35

www.PressureSystems.com

Example:

Supply each of the previously-specified three (3) pressure calibration points to the Multi-

Point Calibration function, as was stated in the previous example of the Configure and

Start (‘00’) sub-command. Assume that all the affected four (4) channels have

differential transducers with the same -5 to +5 psi range. Include at least one pressure

point in the negative range of these transducers

C 01 1 -2.5

C 01 2 0.0

C 01 3 5.0


Read responses (separately after each command executed above):


“-2.4998 -2.4999 -2.5001 -2.500”

“0.0 0.0013 -0.0133 -0.00001”

“5.0091 4.9992 5.0010 4.9998”

Data are returned in reverse channel number order (i.e., 4, 3, 2, 1) in each response.