Pressure Systems, Inc. Model 9116 User’s Manual

Page 30

www.PressureSystems.com

RESET (Command ‘B’)

Purpose: Instructs the module to reset internal operating parameters, and to set all internal

control variables to their default “reset” state (see description below). The current

TCP/IP socket connection will remain open. Execution after a power off/on cycle

is optional (unnecessary).


Command

B

‘B’ is the command letter.


Response

“A”

‘A’ is the acknowledge letter.


Description: The module returns to the following “reset” states if this command is executed:

● Re-zero correction (offset) terms are set to the last values stored in

transducer

memory.

● Span correction (gain) terms are set to the last values stored in transducer

memory.

● Calibration Valve is set to the RUN Position

● Number of Samples for Data Averaging is set to last value stored in non-

volatile memory (factory default = 8).

● Any

autonomous host data delivery streams defined by ‘c’ sub-commands

are

reset

(undefined).

● The

Multi-Point Calibration function defined by ‘C’ sub-commands is reset

(undefined) if in progress.


Example:


Send TCP/IP command to a Model 9116 module (via open socket) to reset defaults:


B


Read following response:


A