Pressure Systems, Inc. Model 9116 User’s Manual

Page 51

www.PressureSystems.com


Example:

Configure three (3) separate autonomous host delivery streams, and divide the module’s

channels between them. Channels (1-4) must be delivered to host as fast as possible,

channels 5-8 may be delivered at half that rate, while the remaining channels 9-16 are

delivered at half the previous rate. All streams are generated continuously and

synchronized with the internal clock at 100 msec., 200 msec., and 400 msec. periods,

respectively. Data are requested in single precision binary IEEE float format f=7).


c 00 1 000F 1 100 7 0

c 00 2 00F0 1 200 7 0"

c 00 3 FF00 1 400 7 0"


Read

response:


“A”

“A”
“A”

To similarly acquire data at “relative” rates (1, 2, and 4) using a periodic hardware
trigger (assumed to also cycle at 10 Hz rate), enter the commands:


c 00 1 000F 0 1 7 0

c 00 2 00F0 0 2 7 0

c 00 3 FF00 0 4 7 0


Read

responses:


A

A

A



Note

The type of data delivered for each specified channel (when the
streams are started) is EU pressure unless sub-command “05” is
also executed to select
other types of data in each stream.