Pressure Systems, Inc. Model 9116 User’s Manual

Page 81

www.PressureSystems.com

cc

Transducer Coefficients Description

Datum

Type

00

Re-zero Cal Adjustment (offset) term

(Note 1) FLOAT

01

Span Cal Adjustment (gain) term

(Note 2) FLOAT

07

User Defined Field

(Note 4) INTEGER

09

Transducer Manufacturing Reference Number (Note 5) INTEGER

0A

Transducer Full-Scale Range Code (See Appendix F) (Note 5) INTEGER


(Note 1)

Related command ‘w08’ can be used to download the offset term to the
sensor’s non-volatile memory (digitally or non-digitally compensated sensor).


(Note 2)

Related command ‘w09’ can be used to download the gain term to the
sensor’s non-volatile memory (digitally or non-digitally compensated sensor).


(Note 4)

Data is immediately stored to the sensor’s non-volatile memory.


The

User Defined Date field (cc=07) is a 32-bit integer. Possible uses are to I

ndicate the date of last user zero and/or span calibration or possibly the date of
next required calibration. If this optional field is used, the user is responsible for
correctly encoding the date into the appropriate 32-bit integer value e.g., a
decimally encoded ‘yymmdd’ date. Any modifications of this field (using the
Download Internal Coefficients (‘v’) command) will result in the new value
automatically being entered to transducer non-volatile memory.


A

special

single

Other Coefficients array is selected with array index aa=11

(hex). All the valid coefficient indexes (for this array only) are listed in the

following

table:


cc

Other Coefficients Description

Datum

Type

01 EU

Pressure Conversion scaler (default=1.0)

FLOAT