Measurement Specialties 9116 User Manual

Page 76

Advertising
background image

Pressure Systems, Inc. Model 9116 User’s Manual

Page 67

www.PressureSystems.com

READ TEMPERATURE VOLTAGES (Command ‘n’)

Purpose: Returns the most recently acquired raw temperature data for the specified

channels converted to engineering-unit Volts directly from the averaged A/D

counts. It is similar to command ‘V,’ except that the raw data reflects a channel’s

temperature signal instead of its pressure signals. Each datum returned in the

response will be in the specified high-precision data format. This command is

intended for advanced users only and is not required for normal operation.


Command

nppppf

n’ is the command letter
pppp’ is the position field
f’ is the format field


Response

dddd.. dddd

dddd’ are the datum fields, each with a leading space (except f= 7
or 8).


Description: The 4-character hex position field (pppp) specifies a 16-bit binary bit-map, with

each bit (set to 1) specifying a particular channel number (16-1, left-to-right).


The

1-character

format field (f) specifies the format of each datum field (dddd)

that will be returned in the requested response. The first datum returned in the

response will be for the highest channel number supplied, and each (non-binary)

datum will be preceded by a space character. Some formats may not be

applicable to the specific type of data being requested. Valid formats are shown

in the following table:

f

converts each internal response datum value from . . .

max. char.

0 single binary float

to 7-10-digit signed decimal “ [-xxx]x.xxxxxx”

13

1 single binary float

to 8-digit hex integer “ xxxxxxxx”

9

2 double binary float

to 16-digit hex integer “ xxxxxxxxxxxxxxxx”

17

5 single binary float

to long integer (EU*1000) then to 8-digit hex integer

9

7 single binary float

to single binary float (big endian: msb first)

4

8 single binary float

to single binary float (little endian: lsb first)

4


Advertising