Measurement Specialties 9116 User Manual

Page 78

Advertising
background image

Pressure Systems, Inc. Model 9116 User’s Manual

Page 69

www.PressureSystems.com

READ MODULE STATUS (Command ‘q’)

Purpose:

Returns requested module status information.


Command

qii

q’ is the command letter
ii’ is the status index field


Response

hhhh
hhhh’ is a 4-digit hex datum (or other (**) decimal datum


Description: The 2-digit hex index field (ii) chooses a particular status field to be returned.

Returned value is described in the following table for each index (a third column

shows

any ‘w’ command index for setting same option.


ii

returned value

4-digit hex or other decimal (**)

‘w’ set

index

00

Module’s Model Number, as decimal (**) integer value (e.g, 9116).

01

Firmware Version, as hex value
(expressed internally as integer version * 100).
(e.g. hex ‘0100’ = 256 decimal, means Version 2.56)

Power-up Status, as 16-bit hex bit map, bits having the following
meaning:

Bit 0 (LSB):

A/D Failure Error.

Bit 1: Transducer Re-zero Adjustment (offset) Term Range Error (out-

of-range values set to 0.0 internally).

Bit 2: Transducer Span Adjustment (gain) Term Range Error (out-of-

range values set to 1.0 internally).

Bit 3: Temperature Correction Coefficients Not Present or Out-of-Range

(if transducer has one or more bad coefficients, all set to 0.0).

Bit 4: reserved (for transducer checksum)

Bit 5: FLASH Initialized Data Section Checksum Error (if error, all data

variables set to factory defaults and stored in FLASH).

02

Bit 6: SRAM Error.

03 reserved

04 reserved

05

Number of A/D Samples To Average, as hex value (e.g., 000A=10
decimal).

10

Advertising