Pressure Systems, Inc. Model 9116 User’s Manual

Page 108

www.PressureSystems.com

5.1.3.3

Remove and Replace PC-323 (Daughter board) on PC-322 (Main PCB)


(1)

Remove the two (2) Phillips-head screws that hold the PC-323 onto the PC-322 board.

(Figure 5.3a)




(2)

Gently rock the PC-323 board back and forth to loosen it and then lift straight up to

remove it. Place the old PC-323 in an electrostatically-protected bag for possible repair

at the PSI factory.


(3)

Remove the new PC-323 board from its electrostatically-protected container. Ensure the

nylon spacer is in place on top of the mounting bar and over the threaded hole. (See

above

photograph,

Figure 5.3b). Align the two (2) 40-pin connectors and press the

board into place.

Remove Phillips-head screws

(screw already removed)

Figure 5.3a

Removing PC-323 Daughter Board

Nylon spacer

Figure 5.3b

PC-323 Daughter Board Removed froom PC-322