Pressure Systems, Inc. Model 9116 User’s Manual

Page 118

www.PressureSystems.com


Chapter 6


Troubleshooting Guide

6.1

Ethernet Module Troubleshooting

6.1.1

Checking Module Power-Up Sequence


(1) Proper power to the module should first be verified. If possible, verify that the output of

the module power supply is set within the range of 18-36 VDC. This should be nominally

set for 24 VDC. Ensure the power supply setting is high enough to compensate for
cable voltage drops if long interface cable lengths are used.


(2)

Turn module power switch ON and verify the following top panel LED status following

initial power-up :

PWR LED should remain ON

If this LED is not on, all other LED’s will likely also be off. Check the PSI 90DB, remote

power supply (8491), or customer provided power supply to ensure the proper voltage

(18-36 VDC) is being provided. Also verify that the power pins in the module interface

cable are wired as described in Section 2.3.2 and Appendix D

COL LED will illuminate briefly upon power-up. This gives a visual indication that the

LED is functional. Thereafter, the LED should remain OFF.

Tx LED will illuminate briefly upon power-up. This gives a visual indication that the LED

is functional. Subsequent activity on the Tx LED during the power-up

sequence is indication that the RARP/BOOTP protocol is enabled. This will typically

occur following the initial busy (BSY) LED cycle and continue until an appropriate RARP

reply is received.

LNK LED will illuminate briefly upon power-up. This gives a visual indication that the

LED is functional. Thereafter, the LED will indicate proper connection to an Ethernet
hub or switch, and should remain ON.

If this LED is OFF, verify that the module is properly connected to the

communications hub or switch. Verify proper power is applied to the hub. Also

try connecting the 9116 cable to a different port of the hub. Note that most hubs

have similar link LEDs to indicate proper connection to the hub itself. If present,

verify that the hub link LED for the pressure scanner and the host computer are

both active. If the hub is functioning correctly, verify that the communications

pins in the module interface cable are wired as described in Section 2.3.4.1 and

Appendix D.