Measurement Specialties 9116 User Manual

Page 72

Advertising
background image

Pressure Systems, Inc. Model 9116 User’s Manual

Page 63

www.PressureSystems.com

CALCULATE AND SET OFFSETS (Command ‘h’)

Purpose:

Instructs a module to calculate new offset coefficients with zero differential

pressure

(or

a

specified “generated” pressure) applied to the specified channels.

These new coefficients update part of the module’s internal calibration coefficient

database, used to convert any subsequent raw data into engineering units data.

The new offset values are also returned in the response. This command is

sometimes called a Re-zero or zero-only calibration.


Command

hpppp [vv.vvvv]

h’ is the command letter
pppp’ is the position field
vv.vvvv
is an [optional] applied pressure value preceded by a space
character


Response

g.gggg .. [g.gggg]...”

g.gggg’ are the actual offset data values returned, each preceded by a
space.



Description: The position field may have 0 or 4 characters. If no position field is specified,

offset coefficients for all of a module’s input channels will be calculated and

returned. If a position field is specified, offset coefficients for only the channels

whose bits are set (=1) will be calculated and returned. If the optional pressure

value

[

vv.vvvv] is specified, the position field must be 4 characters, even when

all channels are to be specified. If the optional pressure value [‘vv.vvv’] is not

provided, an applied pressure of 0.0 psi(a) will be assumed when calculating

coefficients. Offset values are returned in the response in order of highest

specified

channel to lowest specified channel, with data formatted per an implied

decimal format (f=0).

Before acquiring data with this command, any addressed Model 9116 module
will

normally attempt to place the calibration valve in the CAL position, so that a

zero differential pressure can be applied to all channels via the module’s CAL
and CAL Ref input port. Simply leaving these ports unattached will allow the
transducers to read the appropriate zero differential pressure if ambient air
pressure is stable. After data are acquired, the calibration valve will be placed in
the RUN position. To disable the automatic shifting of the calibration valve, refer
to

the

Set Operating Options (‘w’) command (index 0B). The reader is also

referred

to

Section 4.2 of Chapter 4 for additional details concerning the

performance of a Re-zero Calibration.

Note

The calculated offset values from the latest ‘h’ command will be lost
when the module is powered off. To save these offset terms to each
transducer’s non-volatile memory refer to the Set Operating Options

(‘w’) command (index 08).

Advertising