Pressure Systems, Inc. Model 9116 User’s Manual

Page 83

www.PressureSystems.com

SET/DO OPERATING OPTIONS/FUNCTIONS (Command ‘w’)


Purpose:

Change a module’s default operating option settings, or invoke special internal

operations or functions.


Command

wii[dd[ eeee]]

‘w’ is the command letter.
ii’ is the option index field.
dd’ is an [optional] hex datum field.
eeee’ is an [optional] extra datum with a leading space character.


Response

A
‘A’ is the acknowledge letter



Description: The index field (ii) contains two hex digits that identify the specific option to be

set or function to be performed. The datum field (dd), when present, contains 2

hex digits. A few indexes also require an extra datum field (eeee). Valid

options/functions are listed in the table below (-- marks a missing datum field in

its column, and fourth column shows any ‘q’ command index that reads same

option):


ii

dd

Description

‘q’ read

index

00 ---

Execute Internal Self Test.

01 ---

Update Internal Thermal Coefficients.

02-06

---

Reserved for factory use

07 ---

Store Operating Options in non-volatile flash memory.

08 ---

Store Current Offset Terms in transducers’ non-volatile
memories.

09 ---

Store Current Gain Terms in transducers’ non-volatile
memories.

0A 01-10

Set

Number of Channels in Module (default =16 for 9116).

0B

00


01

Enable Automatic Shifting of Calibration Valve during
Calculate and Set Offsets (‘h’) command (default).
Disable Automatic Shifting of Calibration Valve in ‘h’.
User will manually control calibration value.