Pressure Systems, Inc. Model 9116 User’s Manual

Page 38

www.PressureSystems.com

Command ‘C’— Sub-command Index 03: Abort Multi-Point Calibration

This sub-command has no additional parameters.


Command

C 03

‘C’ is the command letter.
‘ 03’ is the sub-command index (ii) for Abort.

NOTE: all parameters are separated by a space.


Response

A

‘A’ is the acknowledge letter


Description: Aborts the Multi-Point Calibration function, if it is currently in progress. This

sub-command also restores the original “A/D samples for averaging” count to the

module that was in use before the calibration function was started.

It should be noted that executing the Configure & Start (‘00’) sub-command
again, after the calibration function has started collecting data (per Collect Data
(‘01’) sub-commands), but before the final data are calculated (per Calculate &
Apply
(‘02’) sub-command), will have the same affect as this Abort function.

Example:

Abort the Multi-Point Calibration function previously started


C 03


Read

response:


“A”