Measurement Specialties 9116 User Manual

Page 77

Advertising
background image

Pressure Systems, Inc. Model 9116 User’s Manual

Page 68

www.PressureSystems.com

Example:

Send TCP/IP command to Model 9116 module (via its connected socket) that

returns decimal voltage data (of the raw temperature signal) for channels 1, 5, 9,

and

13:


n11110


Response contains data for channels 13, 9, 5, and 1 (left to right):


0.53013 0.541698 0.503633 0.000000


In this example channels 13, 9, and 5 return normal temperature voltage signals

in the range of 0.5 to 0.6 volts. Note that channel 1 returns a value of 0.0 volts,

indicating a possible error in its temperature signal.

Advertising