Pressure Systems, Inc. Model 9116 User’s Manual

Page 86

www.PressureSystems.com

The Valve Position indexes (ii=12 and ii=0C) each have two states (00/01) that when combined
provide four (4) possible states of the C1/C2/C3/C4 internal valves. This “Logical Rotary
Switch” with four (4) positions is summarized in the following chart:

C1 Energized

C2 Not

(0C=01)

C2 Energized

C1 Not

(0C=00)

C3 Energized

C4 Not

(12=01)

PURGE

position

LEAK/CHECK

position

C4 Energized

C3 Not

(12=00)

CAL/RE-ZERO

position

RUN

position



Example:

Send TCP/IP commands to Model 9116 module (via its connected socket)

setting the calibration valve to the CAL (or Re-Zero) position:


w1200" (Set RUN/CAL valve position)

w0C01” (Set

CAL position)


Responses

(both

commands):


“A”

“A”


Note

If the programmer “knows” that the module is already in the RUN/CAL
valve mode, the first command above is optional. For more information,
see Figures 4.1 through 4.4 in Chapter 4.