881/882 Video Test Generator User Guide

(Rev A.22)

643

1. For the 882E analyzer connect the cables as follows:

a. Apply power to the Encryption Status Tester (EST) using the power adapter

supplied with the EST.

b. Connect an HDMI cable from the repeater device under test input to the EST HDMI

port labeled, “HDMI OUT.”

c. Connect an HDMI cable from the 882 HDMI Rx 1 port to the HDMI output of the

repeater device under test.

d. Connect an HDMI cable from the 882 HDMI Tx 1 port to the EST HDMI port labeled

“HDMI IN.”

e. Connect an HDMI cable from the 882 HDMI Tx 2 port to the EST HDMI port labeled

“HDMI.”

The following diagram depicts the test setup for the 882EA:

2. Ensure that the 882 settings used for the 3C-I-0x series described in

To run the 3C-II

series HDCP compliance tests on an HDMI repeater:

” on page 644, are maintained

during this test. This includes selection of the HDMI interface, selection of the key PCP
parameters such as the device type, etc.

3. To run 3C-I-01 tests using the short read with the encryption enable pulse positioned

at the beginning of the vsync pulse, enter the following commands:

CPTX:CRIP 0

// Specifies short read.

CPTX:CPTU

// Initiates the execution of the test.

4. To run 3C-I-01 tests using the combined read with the encryption enable pulse

positioned at the beginning of the vsync pulse, enter the following commands:

CPTX:CRIP 1

// Specifies combined read.

CPTX:CPTU

// Initiates the execution of the test.

Refer to

To view the HDCP compliance report:

” on page 651 for procedures on how to

view the generated report.