Configuring special sync for fs, ls, or cs – Quantum Data 881 User Manual

Page 689

Advertising
background image

881/882 Video Test Generator User Guide

(Rev A.22)

573

Configuring special sync for FS, LS, or CS

The special sync output can provide frame sync, line sync, or composite sync signals. The
output is controlled by setting parameters of these signals using the command line
interface. The following commands are used to set the special sync signal:

# !

- Frame sync

# !

- Probe signal

:# !

- Line sync

%# !

- Composite sync

Note: The special sync output is disabled whenever these parameters are set to zero.

A frame pulse can be generated and output at the beginning of each frame. In the case of
interlaced video, the frame pulse is output at the beginning of the blanking interval (of the
first field) that immediately precedes the top line of active video.

To configure frame sync on the special sync output:

1. Establish a session with the generator using either HyperTerminal over a serial

connection or Telnet over an Ethernet LAN. See

Establishing a terminal session with

the generator

” on page 31 or

Establishing a Telnet session with the generator

” on page

34.

2. Enable the frame sync by entering the following command:

FSPG

1

The values of the other parameters (

:#;

and

%# !

) do not matter once FSPG is

enabled.

FSPG=1, FSPP=0

FSPG=1, FSPP=1

FSPW (pixels)

FSPD (lines)

Active

Blank

Blank

Active

Field 0

Field 1

VSPD (lines)

VID

VS

FS

FS

Advertising
This manual is related to the following products:

882