Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A

Table 10-2: Test Function Values

Test Function

Meaning

Acceptable Values

Corrective Action for
Unacceptable Values

RANGE

The Current Full Scale
Range Setting of the
analyzer's analog outputs

Any None

required

CO MEAS

The most recent detector
reading taken in Measure
mode

2500-4800 mV

Check and adjust IR source and
Sync Demodulator as described
in Section 10.6.5 and 10.6.6

CO REF

The most recent detector
reading taken in
Reference mode

2500 - 4800 mV

Check and adjust IR source and
Sync Demodulator as described
in Section 10.6.5 and 10.6.6

MR RATIO

The ratio of the CO
MEAS value to CO REF
value

1.00 - 1.5

Check CO REF and CO MEAS
values as described above.

PRES

The absolute pressure of
the sample gas in the
absorption cell

0" - 1.0" Hg below
ambient pressure

Check for pneumatic system
problems. See Section 10.6.1.
Check for pressure transducer
problems. See Section 10.6.5

SAMPLE FLOW Sample flow rate

700 - 900 sec/min

Check for pneumatic system
problems. See Section 10.6.1.
Check for flowmeter problems.
See Section 10.6.5

SAMPLE TEMP The temperature of the

sample gas in the
absorption cell

30° - 50°C
(After warm-up)

See Section 10.6.2

BENCH TEMP

The temperature of the
Optical Bench

50°C
(After warm-up)

See Section 10.6.2

WHEEL TEMP

The temperature of the
Gas Filter Wheel

68°C
(After warm-up)

See Section 10.6.2

BOX TEMP

The temperature inside
the analyzer chassis

up to 10°C above
ambient

See Section 10.6.2

DCPS

DC Power Supply
reference - A composite
of all voltages provided
by the DC Power Supply

2500

± 200 mV

See Section 10.6.4

10-6