Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A

Table 5-2: I/O Signals (Continued)

#

Signal

Control

Description

8

WHEEL_HTR

YES

Shows the status of the filter wheel heater. This has
the same function as the LED in the power supply
module.

9

DARK_CAL

YES

Turns off the detector input to the Sync/Demod board
for electronics calibration.

10

ELEC_TEST

YES

Activates the Electric Test diagnostic circuitry

11

BROWNOUT_RESET

YES

Resets the DISP_BROWNOUT circuitry described
above

12

ST_CONC_ALARM_1

YES

Status Bit - gas alarm 1

Logic high = M300M in alarm 1 mode

Logic low = Not in alarm 1 mode

13

ST_CONC_ALARM_2

YES

Status Bit - gas alarm 2

Logic high = M300M in gas alarm 2

Logic low = Not in alarm 2 mode

14

ST_FLOW_ALARM

YES

Status Bit - Flow alarm

Logic High = Sample flow out of spec

Logic Low = Flows within spec

15

ST_TEMP_ALARM

YES

Status Bit - Temperature alarm

Logic High = One or more temps out of spec

Logic Low = Temps within spec

16

ST_DIAG_MODE

YES

Status Bit - In Diagnostic mode

Logic High = M300M in Diagnostic mode

Logic Low = Not in Diag mode

17

ST_POWER_OK

YES

Status Bit - Power OK

Logic High = Instrument power is on

Logic Low = Instrument power is off

18

ST_PRESS_ALARM

YES

Status Bit - Flow alarm

Logic High = Sample pressure out of spec

Logic Low = pressure within spec

(table continued)