Peration, Erification, Tch the table 1-2 values – Teledyne 300M User Manual

Page 24

Advertising
background image

Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A

Table 1-2: Final Test and Calibration Values

Test Values

Observed
Value

Units

Nominal Range

CO MEAS

mV

2500 - 4700

CO REF

mV

2500 - 4700

MR RATIO

1.1 – 1.22

SAMPLE PRESS

in-Hg

27 - 30

VACUUM PRESS

in-Hg

10

± 5

SAMPLE FLOW

CC/MIN

700 - 900

BENCH TEMP

°C 48

± 1

DC POWER
SUPPLY

mV

2500

± 200

SAMPLE TEMP

°C

45-50

BOX TEMP

°C

8- 40

WHEEL TEMP

°C 68

± 2

Diagnostic

Dark Ref

mV

125

± 50

Dark Meas

mV

125

± 50

Electrical test

PPM

400

± 4

Span and CAL Values

CO Span Conc

PPM

400

CO Slope

1.0

± 0.3

CO Offset

-0.1 to +0.1

Noise at Zero (p-p)

PPM

0.5

Noise at Span (p-p)

PPM

<1.0% of reading

Factory Installed Options

Option Installed

Factory Installed Options

Option Installed

Power Voltage/Frequency

Rack Mount, w/ Slides

Internal Zero Span Valves

Analog Voltage Range

0 - V

PROM #

Serial #

Date

Technician

1-16