Table 1-2: Final Test and Calibration Values
Test Values
Observed
Value
Units
Nominal Range
CO MEAS
mV
2500 - 4700
CO REF
mV
2500 - 4700
MR RATIO
1.1 – 1.22
SAMPLE PRESS
in-Hg
27 - 30
VACUUM PRESS
in-Hg
10
± 5
SAMPLE FLOW
CC/MIN
700 - 900
BENCH TEMP
°C 48
± 1
DC POWER
SUPPLY
mV
2500
± 200
SAMPLE TEMP
°C
45-50
BOX TEMP
°C
8- 40
WHEEL TEMP
°C 68
± 2
Diagnostic
Dark Ref
mV
125
± 50
Dark Meas
mV
125
± 50
Electrical test
PPM
400
± 4
Span and CAL Values
CO Span Conc
PPM
400
CO Slope
1.0
± 0.3
CO Offset
-0.1 to +0.1
Noise at Zero (p-p)
PPM
0.5
Noise at Span (p-p)
PPM
<1.0% of reading
Factory Installed Options
Option Installed
Factory Installed Options
Option Installed
Power Voltage/Frequency
Rack Mount, w/ Slides
Internal Zero Span Valves
Analog Voltage Range
0 - V
PROM #
Serial #
Date
Technician
1-16