Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A

Table 1-1: Status Outputs

Output#

Pin Pair
(Low, High)

Definition

Condition

1 1,2

ALARM

1

Gas concentration is exceeding the Hi set
limit.

2

3,4

ALARM 2

Gas concentration is exceeding the HiHi set
limit.

3

5,6

SYSTEM OK

True if no alarm conditions exist

4

7,8

TEMP ALARM

True if any temp warning exists

5

9, 10

DIAG MODE

True if in diagnostic mode

6

11,12

POWER ON

True if main power is on

7

13,14

PRESS ALARM

True on low pressure

8 15,16

HIGH

RANGE

SELECTED

True if the auto-range function has switched
the analyzer into high range.

9

17,18

FLOW ALARM

True if a flow warning exists

10 19,20

RESERVED

11 21,22

SOURCE
WARNING

True if the analyzer source intensity is out
of limits.

12 23,24

RESERVED

1.6.4 RS-232

The RS-232 connection is a male, 9-pin D-sub connector at the location shown in Figure 1-3.
See also Appendix A for additional information.

1.6.5 Pneumatic System

The Model 300M is equipped with a vacuum pump capable of pulling 800 cc/min across a flow
restrictor. This allows a smooth, stable flow of sample through the analyzer.

Sample enters the analyzer through a particulate filter element (47 mm diameter) mounted
immediately behind the front panel. The sample then enters directly into the sample cell. Please
see Figure 1-1 for a flow diagram and Figure 1-4 for pneumatic connections.